Range | 1.0000mOhm to 100.00kOhm | ||||
Resolution | 0.1μOhm | ||||
Over range | 80% to 18,000 | ||||
Max. Uncertainty | 0.03% read. + 2dgt | ||||
Measuring Speed/Time | 10 measurements/second | ||||
Display | 4-½ digits |
Current @ | 1 mOhm | 1 A |
10 mOhm | 1 A | |
100 mOhm | 100 mA | |
1 Ohm | 10 mA | |
10 Ohm | 1 mA | |
100 Ohm | 1 mA | |
1 kOhm | 100 μA | |
10 kOhm | 100 μA | |
100 kOhm | 10 μA |
Measuring Voltage | 4V (20mV limit for contacts) |
Max. Inductive Power | < 100 VA |
Temperature Comp. | With PT100 Probe Range 20°C/23°C |
Temp. coefficient | Off, Cu or Al |
Zero Correction | Compensated at each measurement |
Input Protection | Against inductive spikes |
Error detection display and message (RS232) | Open potential leads: ‘SEN’ (‘ESEN’) |
Open current leads: ‘CUR’ (‘ECUR’) | |
Over range >80%: ‘OVR’ (‘EOVR’) | |
Limits | 0... 18,000 upper/lower |
Limits overrun display and message (RS232) | > or < in front of measured value |
Limit settings | Manual (keys) or RS232 |
Ports/Interface | RS232, Parallel Printer, PLC |
Measuring Start Via | Key, RS232, PLC, Potential free contact |
Dimensions | 280 x 80 x 240 mm (W x H x D) |
Weight | Approx. 3kg |
Environment | -10°C to 40°C |