Micro-Ohmmeter 5897 di động hiện tại cao (100 amp)

An accurate mains/battery powered portable 100A Micro-Ohmmeter. Using the 4 terminal measurement technique the Tinsley 5897 gives basic accuracy of 0.2% and an overall resistance range of 0.1 µOhm to 40 Ohm.
 
With an internal mains re-chargeable battery supply the Tinsley 5897 may be used "in the field" for periods of up to 8 hours without requiring recharging. A standard RS232 computer interface is included that may be used for printing resistance values or downloading measurements that have been stored in the instrument's memory. The Tinsley 5897 Portable Micro-Ohmeter can be used fully auto-ranging or with manual range selection if required. Each of the 6 ranges has a choice of high or low measuring current. Current can be selected to be positive or negative, with the option of automatic reversal to eliminate the effect of thermal E.M.F.'s.
Micro-Ohmmeter 5897 di động hiện tại cao (100 amp)

Tổng quan

Nguyên lý đo lường
Một Kelvin, bốn thiết bị đầu cuối, nguyên tắc đo lường để loại bỏ điện trở chì. Dòng điện 100A được đập với chu kỳ hoạt động khoảng 10%. Giá trị đo hiển thị liên tục.
 
Bộ nguồn
Hai pin axit chì bên trong, có thể sạc lại, được niêm phong, được trang bị cho các mạch cung cấp điện trở thử nghiệm bên ngoài 100A và mạch cung cấp đo lường bên trong. Hai chỉ số tính phí cho thấy trạng thái của phí. Bộ sạc pin tích hợp có nguồn cung cấp đầu vào chính cho bảng điều khiển phía trước để cung cấp 115V hoặc 230V với cầu chì
 
Màn hình kỹ thuật số
Màn hình hiển thị gồm 4 chữ số gồm 4000 chữ số cung cấp độ phân giải tới 0,025% tỷ lệ đọc toàn màn hình trong ánh sáng ban ngày khi xem màn hình LED cao 20 mm
 
Cơ khí
Vỏ nhựa ABS chắc chắn, cầm tay, màu xám. Kích thước 460 x 380 x 160mm. Trọng lượng tịnh xấp xỉ 17kg
 
Hiệu chuẩn
Được cung cấp Giấy chứng nhận hiệu chuẩn Tinsley trong nhà có thể truy nguyên theo Tiêu chuẩn quốc gia thông qua NPL
 

Control Functions

  • 6 measuring range controls (manual selection)
  • AUTO key for auto range
  • High/Low current control
  • 3 Current direction measurement controls
  • + direction
  • - direction
  • Alternate (automatic reversal of + and - to eliminate thermal e.m.f. errors in measuring circuit)
  • Supply ON/OFF
  • Measure key
  • Print key
  • Calibration (CAL) control. Enables instrument internal reference standards to be calibrated by applying a set of external reference standards
  • Menu/Enter key to access additional functions
  • Time and date
  • Print mode
  • Baud rate
  • Data logging mode
  • RS232 for data and printer output
  • 4 large 200A rated input terminals
  • Mains input charging socket
  • 2 battery charge indicators
Specifications

See the table below for a run-down of this product's specifications.

Range
Resolution 
 
Current
Accuracy
Temperature Coefficient
(%reading + last digit)
% reading 
last digits 
400µΩ
0.1µΩ
100A/25A
+/-0.4%
+4
(0.05% +4)/ °C
4000µΩ
1.0µΩ
100A/25A
+/-0.2%
+4
(0.04% +0.2)/ °C
40µΩ
10 µΩ
10A/2.5A
+/-0.2%
+4
(0.02% +0.1)/ °C
400µΩ
0.1µΩ
1A/0.25A
+/-0.2%
+4
(0.02% +0.1)/ °C
4µΩ
1.0µΩ
100mA/25mA
+/-0.2%
+4
(0.02% +0.1)/ °C
40µΩ
10 µΩ
10mA/2.5mA
+/-0.2%
+4
(0.04% +0.2)/ °C